Tibial Baseplate-Cement Interface Debonding in the ATTUNE Total Knee Arthroplasty System

Open AccessPublished:September 20, 2022DOI:https://doi.org/10.1016/j.artd.2022.06.012

      Abstract

      Background

      Acrylic bone cement is the most common method of fixation for primary total knee arthroplasty (TKA). Several studies have described good short-term outcomes; however, there have been reports of early failures due to tibial baseplate debonding at the implant-cement interface of The ATTUNE Knee System (DePuy Synthes, West Chester, PA). We examined the causes and rates of revision in patients who underwent TKA with this system to identify factors associated with this mode of early failure.

      Methods

      A retrospective review of electronic health records between 2013 and 2018 identified all patients undergoing TKA with the ATTUNE Knee System with a minimum 2-year follow-up. Cause of revision, patient, implant, instrumentation, cement, and surgeon variables were collected. A descriptive analysis was used to identify characteristics of surgeon (fellowship-trained, surgical volume), implant (baseplate, bearing), and cement (brand, viscosity) that were associated with aseptic loosening.

      Results

      A total of 668 patients representing 742 knees were identified. Eighteen (2.4%) required a revision surgery. Aseptic loosening was the leading cause of revision surgery (n = 10, 55.6%). All failures due to aseptic loosening involved debonding of the tibial implant-cement interface. A multivariate analysis identified low-volume surgeons (9.0%, P < .0001) and 1 specific brand of high-viscosity cement (14.3%, P < .0001) as risk factors for aseptic loosening.

      Conclusions

      This study represents the largest nonregistry review of the original ATTUNE Knee System. Surgeon case volume and cement viscosity were factors associated with an increased rate of early failure due to tibial baseplate implant-cement interface debonding.

      Keywords

      Introduction

      The volume of total knee arthroplasty (TKA) has dramatically increased over the last decade with 3 million surgeries projected to be performed annually by 2030 in the United States [
      • Kurtz S.M.
      • Ong K.L.
      • Lau E.
      • Bozic K.J.
      Impact of the economic downturn on total joint replacement demand in the United States.
      ]. The most frequent reasons for failure of primary TKA have changed as implant designs, materials, and surgical techniques which have been refined. Early studies implicated polyethylene-wear-associated osteolysis as a major cause for revision, while more recent investigations have identified infection and instability as more frequent causes [
      • Sharkey P.F.
      • Lichstein P.M.
      • Shen C.
      • Tokarski A.T.
      • Parvizi J.
      Why are total knee arthroplasties failing today—has anything changed after 10 years?.
      ,
      • Delanois R.E.
      • Mistry J.B.
      • Gwam C.U.
      • Mohamed N.S.
      • Choksi U.S.
      • Mont M.A.
      Current epidemiology of revision total knee arthroplasty in the United States.
      ].
      Acrylic bone cement is the most common method of fixation for TKA with low rates of early aseptic loosening [
      • Bonutti P.M.
      • Khlopas A.
      • Chughtai M.
      • Cole C.
      • Gwam C.U.
      • Harwin S.F.
      • et al.
      Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface.
      ,
      • Nakama G.Y.
      • Peccin M.S.
      • Almeida G.J.
      • Neto O.D.A.L.
      • Queiroz A.A.
      • Navarro R.D.
      Cemented, cementless or hybrid fixation options in total knee arthroplasty for osteoarthritis and other non-traumatic diseases.
      ]. Implant selection, cementing technique, and cement selection are all factors under surgeon control, which can affect risk of aseptic loosening. The original ATTUNE total knee system (DePuy Synthes, West Chester, PA) was introduced in 2012 with features aimed at addressing cement adhesion and aseptic loosening. The underside of the tibial baseplate was designed with a grit-blasted surface finish roughness of 2.5-3.5 Ra to better accommodate fixation of bone cement [,]. Early studies demonstrated good short-term outcomes [
      • Ranawat C.S.
      • White P.B.
      • West S.
      • Ranawat A.S.
      Clinical and radiographic results of Attune and PFC sigma knee designs at 2-year follow-up: a prospective matched-pair analysis.
      ,
      • Turgeon T.R.
      • Gascoyne T.C.
      • Laende E.K.
      • Dunbar M.J.
      • Bohm E.R.
      • Richardson C.G.
      The assessment of the stability of the tibial component of a novel knee arthroplasty system using radiostereometric analysis.
      ]. However, there have been reports of early failures due to aseptic loosening of the tibial component, specifically at the implant-cement interface [
      • Bonutti P.M.
      • Khlopas A.
      • Chughtai M.
      • Cole C.
      • Gwam C.U.
      • Harwin S.F.
      • et al.
      Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface.
      ,
      • Cerquiglini A.
      • Henckel J.
      • Hothi H.
      • Allen P.
      • Lewis J.
      • Eskelinen A.
      • et al.
      Analysis of the Attune tibial tray backside: a comparative retrieval study.
      ]. This system was subsequently redesigned in 2017 as seen in Figure 1, introducing a new tibial baseplate, which was reconfigured with an undercut cement pocket area and a greater surface roughness (3.0-6.5 Ra) to enhance cement bonding. Cement debonding of the tibial component during revision surgery was observed independently by 2 of the adult reconstruction fellowship-trained authors (D.K. and C.D.) prompting investigation of the cases and rates of failure (Fig. 2).
      Figure thumbnail gr1
      Figure 1Images designated A demonstrate the 2012 original ATTUNE tibial base plate design compared to images labeled B of the 2017 new S+ design. The lower right panel depicts the described added undersurface “under pocket” features to provide a macrolock at cement-implant interface []. Specifically, the new S+ design was manufactured with greater surface roughness (3.0-6.5 Ra) to enhance cement bonding. The image is reproduced with permission from DePuy.
      Figure thumbnail gr2
      Figure 2Example of the original ATTUNE tibial component retrieved during revision surgery. The undersurface of the component is shown demonstrating the surface completely devoid of cement adhesion. A corresponding intraoperative photograph depicting remaining cement mantle in tibia.
      Currently, the majority of clinical data on the survivorship of this total knee system stem from registry data [,
      • Bonutti P.
      Response to: confidence in the ATTUNE Knee is Driven by Real-World Scientific Evidence: response to Bonutti et al. Article.
      ,
      National Joint Registry for England, Wales
      Northern Ireland and the Isle of Man. Implant summary report for DePuy ATTUNE CR and ATTUNE PS. NJR Database extracted April 7, 2017:1–17.
      ]. Previous studies have failed to establish causes for this mode of failure and were often limited by small sample sizes or brief follow-up [
      • Bonutti P.M.
      • Khlopas A.
      • Chughtai M.
      • Cole C.
      • Gwam C.U.
      • Harwin S.F.
      • et al.
      Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface.
      ,
      • Kopinski J.E.
      • Aggarwal A.
      • Nunley R.M.
      • Barrack R.L.
      • Nam D.
      Failure at the tibial cement-implant interface with the use of high-viscosity cement in total knee arthroplasty.
      ,
      • Dwyer K.
      • Jones R.E.
      • Lesko J.
      • Leopold J.
      • Diaz R.
      ,
      • Kaptein B.L.
      • den Hollander P.
      • Thomassen B.
      • Fiocco M.
      • Nelissen R.G.
      A randomized controlled trial comparing tibial migration of the ATTUNE cemented cruciate-retaining knee prosthesis with the PFC-sigma design.
      ,
      • Sadauskas A.
      • Engh III, C.
      • Mehta M.
      • Levine B.
      Implant interface debonding after total knee arthroplasty: a new cause for concern?.
      ,
      • Lachiewicz P.F.
      • Vovos T.J.
      • Steele J.R.
      • Wellman S.S.
      Cemented ATTUNE® fixed- bearing modular posterior-stabilized knee arthroplasty has an unexpectedly high rate of revision.
      ]. The purpose of our study was to describe the causes and rates of early revision in patients undergoing primary TKA with the ATTUNE total knee system and to identify surgical factors associated with aseptic tibial loosening. We hypothesized that aseptic loosening would be the leading cause of early revision and failure rates, specifically debonding at the tibial implant-cement interface. We also hypothesized a low surgery volume would lead to an increased early revision rate.

      Material and methods

      Institutional review board approval was obtained for this retrospective study. The electronic health record of a large, rural integrated health system between March 1, 2013, and November 15, 2018, was queried to identify patients who underwent cemented primary TKA utilizing the ATTUNE total system (Fig. 2). The use of this TKA system as well as cement type in the primary setting was surgeon preference in all cases and not based on institutional standard or specific patient factor. Patients were included if they had at least 2-year clinical follow-up or revision surgery at an earlier timepoint. Patients were excluded if 2-year follow-up was not available. If patients had inadequate follow-up, they were contacted by phone utilizing an institutional review board-approved script of simple yes or no questions to specifically collect interval history of any information regarding detail on status and date of reoperation or revision. The phone call did not inquire about patient-reported outcomes or complaints and only asked about any surgery or procedure to the knee in question. Three attempts were made to reach the patient, and if unsuccessful, patients were excluded. Patient demographic characteristics, including age, body mass index (BMI), gender, employment status, tobacco use, and history of diabetes, were collected (Table 1).
      Table 1Patient demographics.
      Patient demographicsAllWithout revisionWith revision
      Age in years, mean (SD)70.3 (9.8)70.5 (9.6)61.4 (11.0)
      BMI, mean (SD)33.0 (3.4)33.0 (6.4)35.3 (5.1)
      Male, n (%)260 (38.9)254 (38.9)8 (44.4)
      White, n (%)653 (97.8)636 (97.4)17 (94.4)
      Employed, n (%)116 (17.4)109 (16.3)7 (38.9)
      Active tobacco use, n (%)42 (6.3)38 (5.8)4 (22.2)
      Diabetes, n (%)153 (22.9)150 (23.0)3 (16.7)
      Follow up in years, mean (SD)3.5 (1.4)3.5 (1.4)3.2 (0.9)
      Complications were identified using International Classification of Diseases (ICD-10: T84.∗) diagnosis codes, and all cases were manually reviewed by coauthors (D.T. and B.F.) (ICD-10: international statistical classification of disease and related health problems: tenth revision, second edition, World Health Organization). Cause of revision was determined by review of the electronic health record including office visit documentation and operative reports. Surgeons were categorized based on TKA volume (high >50 vs low ≤50 cases per year) and completion of an adult reconstruction fellowship (yes or no) [
      • Lau R.L.
      • Perruccio A.V.
      • Gandhi R.
      • Mahomed N.N.
      The role of surgeon volume on patient outcome in total knee arthroplasty: a systematic review of the literature.
      ]. Cement was identified by manufacturer (DePuy Synthes, West Chester, PA; DJO Global, Vista, CA; Stryker, Kalamazoo, MI; and Zimmer Biomet, Warsaw, IN) and categorized into viscosity levels (high, medium, or low) based on the manufacturer classification. Only 1 of the 4 manufacturers (DePuy) was associated with 2 different cement viscosity types (medium and high); otherwise, DJO and Zimmer cements were of high viscosity, and Stryker was of low viscosity. Tibial baseplates were classified according to bearing type (fixed or rotating platform) and baseplate type.

      Statistical analysis

      We examined each of the 6 characteristics described above (surgeon volume, surgeon training, cement manufacturer, cement viscosity, bearing type, and baseplate type) individually and used chi-squared or Fisher’s exact tests to compare percentages of cases with revision and aseptic loosening within each. We then divided all cases into all 120 possible combinations of the 6 characteristics and descriptively examined which combinations of factors had the highest percentage of aseptic loosening. A statistical analysis was performed using the SAS software (SAS 9.4; SAS Institute, Cary, NC), with contrasts of P < .05 considered statistically significant.

      Results

      Of 836 cases identified, 742 (89%) met the inclusion criteria (Fig. 3). If patients had inadequate follow-up, they were contacted by phone to collect interval history pertaining to reoperation or revision data to establish the minimum follow-up. Ninety-two out of 836 (11%) were not able to be contacted to establish follow-up. None of these patients had a revision surgery within our system based on manual review of the electronic medical record. Overall, 18 patients underwent a revision surgery, resulting in a 2-year revision rate of 2.4% (95% confidence interval [CI] = 1.4% to 3.8%). Aseptic loosening was the leading cause of revision (n = 10, 55.6% of revisions) for a 2-year aseptic loosening revision rate of 1.3% (95% CI = 0.6% to 2.5%) (Table 2). All cases of aseptic loosening demonstrated debonding at the tibial implant-cement interface (Fig. 2). The mean time to revision was 22.6 months (range: 9.8-36.3 months).
      Table 2Causes of revision.
      Cause of revisionn (%)
      Aseptic loosening10 (55.6)
      Infection3 (16.7)
      Instability2 (11.1)
      Patella maltracking1 (5.6)
      Patella resurfacing1 (5.6)
      Fracture1 (5.6)
      Total18
      Surgeries were performed by 8 surgeons using 5 different types of cement (Table 3). Taken individually, the 2 factors that were significantly associated with all-cause revision were cement manufacturer (17.1% DJO Surgical high-viscosity cement [HVC] vs 0.0%-2.2% for all others, P = .0003) and surgeon volume (10.5% low-volume surgeons vs 1.6% high-volume surgeons, P = .0005) (Table 4). These same 2 individual factors were also significantly associated with higher rates of revision due to aseptic loosening (14.3% DJO Surgical HVC vs 0.0%-1.1% in all others, P < .0001; 9% for low-volume surgeons vs 0.6% high-volume surgeons, P < .0001) (Table 5). Non–fellowship-trained surgeons demonstrated a marginally significant association with higher risk of aseptic loosening (2.1% vs 0.3% for fellowship-trained, P = .05). None of the aseptic loosening cases used low-viscosity cement (LVC) (Stryker), medium-viscosity cement (DePuy), the newer baseplate design, or the revision tibial baseplate design.
      Table 3Surgeon demographics and cement use.
      Surgeon ## Cases contributedFellowship trainedHigh volumeCement brands usedViscosities used
      144NNDJO Surgical, ZimmerHigh
      260NYDePuyMedium, high
      334NYZimmer, Stryker, DePuyLow, medium, high
      423NYDePuyMedium, high
      5307YYZimmerHigh
      6192NYStryker, DepuyLow, medium
      723NNZimmer, StrykerLow, high
      859NYDePuyMedium, high
      N, no; Y, yes.
      Table 4Numbers and percent of patients in each category of risk factor that required revision surgery (for any reason).
      N (%) with revision for any reason (n = 18)P value
      Cement brand, N (%).0003
       DJO Surgical (n = 35)6 (17.1%)
       DePuy (n = 265)4 (1.5%)
       Stryker (n = 74)0 (0%)
       Zimmer (n = 368)8 (2.2%)
      Cement viscosity, N (%).12
       Low (n = 74)0 (0%)
       Medium (n = 182)2 (1.1%)
       High (n = 483)16 (3.3%)
      Surgeon fellowship trained, N (%).24
       Yes (n = 307)5 (1.6%)
       No (n = 435)13 (3.0%)
      Surgeon with high volume, N (%).0005
       Yes (n = 675)11 (1.6%)
       No (n = 67)7 (10.5%)
      Baseplate design, N (%).44
       ATTUNE (n = 615)17 (2.8%)
       ATTUNE S+ (n = 120)1 (0.8%)
       ATTUNE Revision Tibia (n = 7)0 (0%)
      Bearing type, N (%).66
       Rotating platform (n = 63)2 (3.2%)
       Fixed bearing (n = 679)16 (2.4%)
      Table 5Characteristics of patients who underwent revision surgery for aseptic loosening.
      Aseptic loosening (n = 10)P value
      Cement brand, N (%)<.0001
       DJO Surgical (n = 35)5 (14.3%)
       DePuy (n = 265)1 (0.4%)
       Stryker (n = 74)0 (0%)
       Zimmer (n = 368)4 (1.1%)
      Cement viscosity, N (%).07
       Low (n = 74)0 (0%)
       Medium (n = 182)0 (0%)
       High (n = 486)10 (2.1%)
      Surgeon fellowship trained, N (%).05
       Yes (n = 307)1 (0.3%)
       No (n = 435)9 (2.1%)
      Surgeon with high volume, N (%)<.0001
       Yes (n = 675)4 (0.6%)
       No (n = 67)6 (9.0%)
      Baseplate design, N (%).44
       ATTUNE (n = 615)10 (1.6%)
       ATTUNE S+ (n = 120)0 (0%)
       ATTUNE Revision Tibia (n = 7)0 (0%)
      Bearing type, N (%).59
       Rotating platform (n = 63)1 (1.6%)
       Fixed bearing (n = 679)9 (1.3%)
      Of the 120 possible combinations of the 6 characteristics we examined, only 19 were observed as cases in the data, and of those, 9 combinations had 10 or more surgical cases in their subgroup. As summarized in Table 6, the group with the highest percentage of aseptic loosening was non–fellowship-trained surgeons using DJO Surgical HVC (n = 31, 5/31 cases, or 16.1% with aseptic loosening; 95% CI = 5.4% to 33.7%). All surgeons in this group had a low volume (<50 cases per year), and all cases used fixed-bearing tibial implants with the original ATTUNE baseplate design, making it impossible to further isolate risk factors in this group. The combination that led to the second-highest rate of aseptic loosening occurred with non–fellowship-trained surgeons using Zimmer HVC and the original tibial baseplate design (n = 51, 3/51 cases, or 5.9% with aseptic loosening; 95% CI = 1.2% to 16.2%). All implants in this group used a fixed-bearing baseplate, and the cases with low-volume surgeons in this group had a slightly lower percentage of aseptic loosening (1/19 cases, or 5.3%) compared to 2 of 32 cases or 6.3% for the high-volume surgeons. When comparing these 2 groups to the remainder of the cohort, chi-squared testing concluded that the rates of aseptic loosening in both groups were statistically significantly higher than those of the remainder of the population (P < .01).
      Table 6All cases divided into 3 combination categories based on fellowship status and cement used vs all other cases.
      N, % of cases with aseptic loosening [95% CI]P value
      Category<.01
       Non-fellowship-trained, DJO cement (n = 31)5 (16.1%) [5.4%, 33.7%]
       Non-fellowship-trained, Zimmer cement (n = 51)3 (5.9%) [1.2%, 16.2%]
       All other cases (n = 660)2 (0.3%) [0.04%, 1.1%]

      Discussion

      Aseptic loosening at the tibial implant-cement interface is an uncommon mode of early failure but has been reported when using the original ATTUNE total knee system [
      • Bonutti P.M.
      • Khlopas A.
      • Chughtai M.
      • Cole C.
      • Gwam C.U.
      • Harwin S.F.
      • et al.
      Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface.
      ]. This study represents the largest nonregistry review of this contemporary total knee implant with at least 2-year follow-up and is the first to present revision rates. Aseptic loosening was the most common cause for early revision, accounting for over half of early revisions. Interestingly, all cases demonstrated debonding at the tibial implant-cement interface (Fig. 2). Non–arthroplasty-fellowship-trained surgeons using DJO (medium viscosity) or Zimmer (high viscosity) cements were at the highest risk of aseptic loosening with tibial baseplate cement debonding.
      Shortly after the release of this implant system, the manufacturer published its surgical technique guide, describing the design and surface finish on the underside of the tibial baseplate aimed at addressing aseptic loosening and based on internal research []. This design feature has given rise to conflicting outcome reports. Registry data showed low rates of overall revision, and several studies reported good short-term outcomes [
      • Bonutti P.
      Response to: confidence in the ATTUNE Knee is Driven by Real-World Scientific Evidence: response to Bonutti et al. Article.
      ,
      National Joint Registry for England, Wales
      Northern Ireland and the Isle of Man, 13th annual report 2016; Table 3.28.
      ,
      Australian orthopaedic association national joint replacement registry annual report 2016. Tables KT9 and KT22.
      ,
      • Giaretta S.
      • Berti M.
      • Micheloni G.M.
      • Ceccato A.
      • Marangoni F.
      • Momoli A.
      Early experience with the ATTUNE total knee replacement system.
      ,
      • Molloy I.B.
      • Keeney B.J.
      • Sparks M.B.
      • Paddock N.G.
      • Koenig K.M.
      • Moschetti W.E.
      • et al.
      Short term patient outcomes after total knee arthroplasty: does the implant matter?.
      ]. However, early aseptic loosening was described in several studies, and retrieval studies noted poor cement adhesion to the tibial baseplate [
      • Bonutti P.M.
      • Khlopas A.
      • Chughtai M.
      • Cole C.
      • Gwam C.U.
      • Harwin S.F.
      • et al.
      Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface.
      ,
      • Cerquiglini A.
      • Henckel J.
      • Hothi H.
      • Allen P.
      • Lewis J.
      • Eskelinen A.
      • et al.
      Analysis of the Attune tibial tray backside: a comparative retrieval study.
      ,
      • Lachiewicz P.F.
      • Vovos T.J.
      • Steele J.R.
      • Wellman S.S.
      Cemented ATTUNE® fixed- bearing modular posterior-stabilized knee arthroplasty has an unexpectedly high rate of revision.
      ,
      • Kelly B.C.
      • Owen J.R.
      • Shah S.C.
      • Johnson A.J.
      • Golladay G.J.
      • Kates S.L.
      A biomechanical comparison of the effect of baseplate design and bone marrow fat infiltration on tibial baseplate pullout strength.
      ,
      • Staats K.
      • Wannmacher T.
      • Weihs V.
      • Koller U.
      • Kubista B.
      • Windhager R.
      Modern cemented total knee arthroplasty design shows a higher incidence of radiolucent lines compared to its predecessor.
      ]. The revision rate reported here (2.4%) is significantly higher than the 1.3% 4-year revision rate from the National Registry for England, Wales, Norther Ireland, and the Isle of Man, as well as the revision rates for the cruciate retaining option (0.5%) and posterior stabilized option (0.4%) from the 2016 Australian Orthopedic Association National Joints Replacement Registry [
      • Bonutti P.
      Response to: confidence in the ATTUNE Knee is Driven by Real-World Scientific Evidence: response to Bonutti et al. Article.
      ,
      National Joint Registry for England, Wales
      Northern Ireland and the Isle of Man, 13th annual report 2016; Table 3.28.
      ]. All our cases involving aseptic loosening occurred with the original baseplate design and demonstrated debonding at the tibial implant-cement interface (Fig. 1). This mechanism of failure matches previous studies showing poor cement adhesion to the tibial baseplate with this implant design [
      • Bonutti P.M.
      • Khlopas A.
      • Chughtai M.
      • Cole C.
      • Gwam C.U.
      • Harwin S.F.
      • et al.
      Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface.
      ,
      • Cerquiglini A.
      • Henckel J.
      • Hothi H.
      • Allen P.
      • Lewis J.
      • Eskelinen A.
      • et al.
      Analysis of the Attune tibial tray backside: a comparative retrieval study.
      ,
      • Kelly B.C.
      • Owen J.R.
      • Shah S.C.
      • Johnson A.J.
      • Golladay G.J.
      • Kates S.L.
      A biomechanical comparison of the effect of baseplate design and bone marrow fat infiltration on tibial baseplate pullout strength.
      ].
      Numerous factors including BMI, bone quality, component alignment, implant design, and perhaps most importantly, surgical technique contribute to aseptic loosening following TKA [
      • Kelly B.C.
      • Owen J.R.
      • Shah S.C.
      • Johnson A.J.
      • Golladay G.J.
      • Kates S.L.
      A biomechanical comparison of the effect of baseplate design and bone marrow fat infiltration on tibial baseplate pullout strength.
      ]. One potential technique factor that has received recent attention is the viscosity of the cement used during prosthesis implantation. HVC appeals to arthroplasty surgeons due to its fast mixing and waiting phases with concomitant prolonged working and hardening phases compared to LVC [
      • Bonutti P.M.
      • Khlopas A.
      • Chughtai M.
      • Cole C.
      • Gwam C.U.
      • Harwin S.F.
      • et al.
      Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface.
      ,
      • Nakama G.Y.
      • Peccin M.S.
      • Almeida G.J.
      • Neto O.D.A.L.
      • Queiroz A.A.
      • Navarro R.D.
      Cemented, cementless or hybrid fixation options in total knee arthroplasty for osteoarthritis and other non-traumatic diseases.
      ,
      • Cerquiglini A.
      • Henckel J.
      • Hothi H.
      • Allen P.
      • Lewis J.
      • Eskelinen A.
      • et al.
      Analysis of the Attune tibial tray backside: a comparative retrieval study.
      ]. However, prior investigations showed HVC to have less penetration into bone and weaker cement-implant shear strength in a bovine cancellous bone model and a cadaveric femur study 4/5. Recently, a previously rare mode of early aseptic loosening was reported in which the tibial component debonded at the implant-cement interface [
      • Bonutti P.M.
      • Khlopas A.
      • Chughtai M.
      • Cole C.
      • Gwam C.U.
      • Harwin S.F.
      • et al.
      Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface.
      ,
      • Nakama G.Y.
      • Peccin M.S.
      • Almeida G.J.
      • Neto O.D.A.L.
      • Queiroz A.A.
      • Navarro R.D.
      Cemented, cementless or hybrid fixation options in total knee arthroplasty for osteoarthritis and other non-traumatic diseases.
      ]. The use of HVC was implicated as a potential factor contributing to this failure mechanism, even though the vast majority of TKAs performed with HVC have performed well. In our study, DJO HVC was significantly associated with risk of revision and aseptic loosening. Additionally, no cases of aseptic loosening occurred with LVC. Previous studies have also demonstrated higher failure rates with HVC [
      • Lau R.L.
      • Perruccio A.V.
      • Gandhi R.
      • Mahomed N.N.
      The role of surgeon volume on patient outcome in total knee arthroplasty: a systematic review of the literature.
      ,
      • Hazelwood K.J.
      • O’Rourke M.
      • Stamos V.P.
      • McMillan R.D.
      • Beigler D.
      • Robb III, W.J.
      Case series report: early cement–implant interface fixation failure in total knee replacement.
      ,
      • Buller L.T.
      • Rao V.
      • Chiu Y.F.
      • Nam D.
      • McLawhorn A.S.
      Primary total knee arthroplasty performed using high-viscosity cement is associated with higher odds of revision for aseptic loosening.
      ]. The case series by Hazelwood et al. reported on 9 early cases of aseptic loosening of the tibial component [
      • Hazelwood K.J.
      • O’Rourke M.
      • Stamos V.P.
      • McMillan R.D.
      • Beigler D.
      • Robb III, W.J.
      Case series report: early cement–implant interface fixation failure in total knee replacement.
      ]. HVC was used in all 9 cases, and 7 loose components were from a similar implant knee system from the same manufacturer. HVC may be a contributing factor to loosening, but our review identified no cases of femoral component loosening, suggesting the unique repeated failure of the tibial component is feasibly related to tibial component design issues rather than surgeon technique.
      Low-volume surgeons were found to be at a higher risk of revision for aseptic loosening. This association may be related to cementing technique. However, this hypothesis is not proven from our results and is not fully defined in the literature. A recent multicenter review of 4 academic hospitals found low-volume surgeons (<50 cases/year) to be a risk factor for implant malalignment in primary TKA [
      • Kazarian G.S.
      • Lawrie C.M.
      • Barrack T.N.
      • Donaldson M.J.
      • Miller G.M.
      • Haddad F.S.
      • et al.
      The impact of surgeon volume and training status on implant alignment in total knee arthroplasty.
      ]. However, a review of the 64,017 primary TKA cases from a Total Joint Replacement Registry demonstrated that surgeon volume (<10 cases/year, 10-49 cases/year, or ≥50 cases/year) and surgeon fellowship training were not associated with risk of aseptic revision [
      • Namba R.S.
      • Cafri G.
      • Khatod M.
      • Inacio M.C.
      • Brox T.W.
      • Paxton E.W.
      Risk factors for total knee arthroplasty aseptic revision.
      ]. Martin et al. presented a cadaveric study at the American Association of Hip and Knee Surgeons Annual Meeting 2021 after studying if motion during cementing significantly altered tibial implant fixation strength in 4 contemporary TKA designs [
      • Martin R.
      • Mason B.
      • Wronski P.
      • Schilkowsky R.
      • Orfanos A.
      • Fehring T.
      • et al.
      Motion during cementing significantly decreases tibial implant fixation strength.
      ]. They found that knee motion during cement polymerization is associated with significant decreases in tibial implant fixation strength in each implant design and recommended limiting motion while cementing the tibial implant to improve fixation strength [
      • Martin R.
      • Mason B.
      • Wronski P.
      • Schilkowsky R.
      • Orfanos A.
      • Fehring T.
      • et al.
      Motion during cementing significantly decreases tibial implant fixation strength.
      ]. The role of surgeon technique in implant failure remains unclear and is confounded in our review as 1 low-volume surgeon exclusively used HVC.
      Recent literature has shown that the risk for revision TKA due to aseptic tibial component failure is almost 2 times greater in those with BMI ≥35 kg/m2 and independent of age and coronal alignment [
      • Abdel M.P.
      • Bonadurer III, G.F.
      • Jennings M.T.
      • Hanssen A.D.
      Increased aseptic tibial failures in patients with a BMI > 35 and well-aligned total knee arthroplasties.
      ]. The average BMI for our cohort was 34 kg/m2, with 50% of the patients having a BMI ≥35 kg/m2 consistent with other series [
      • Pfefferle K.J.
      • Gil K.M.
      • Fening S.D.
      • Dilisio M.F.
      Validation study of a pooled electronic healthcare database: the effect of obesity on the revision rate of total knee arthroplasty.
      ,
      • Dowsey M.
      • Choong P.M.
      Obese diabetic patients are at substantial risk for deep infection after primary TKA.
      ,
      • Stickles B.
      • Phillips L.
      • Brox W.T.
      • Owens B.
      • Lanzer W.L.
      Defining the relationship between obesity and total joint arthroplasty.
      ]. Martin et al. presented a retrospective cohort of 216 patients revised for aseptic tibial loosening at the American Association of Hip and Knee Surgeons Annual Meeting 2021 in an attempt to determine if there is a difference in cement mantle thickness based on the failure interface [

      Cox Z., Engstrom S., Shinar A., Polkowski G., Mason B., Martin R., et al. Is cement mantle thickness a primary cause of aseptic loosening following primary total knee arthroplasty? TheKNE-D-22-00258. SSRN: https://ssrn.com/abstract=4125881 or http://dx.doi.org/10.2139/ssrn.4125881.

      ]. In their study, 203 patients demonstrated radiographic failure at the implant-cement interface, and 13 patients demonstrated failure at the cement-bone interface. The average cement mantle thickness for all anterior posterior and lateral zones was significantly greater for patients that had failure at the implant-cement interface vs for those with failure at the cement-bone interface (anterior posterior: 4.6 mm vs 1.4 mm [P < .001]; lateral 4.6 mm vs 1.9 mm [P < .001]) suggesting methods for decreasing tibial implant loosening should likely focus on improving the fixation at the implant-cement interface [

      Cox Z., Engstrom S., Shinar A., Polkowski G., Mason B., Martin R., et al. Is cement mantle thickness a primary cause of aseptic loosening following primary total knee arthroplasty? TheKNE-D-22-00258. SSRN: https://ssrn.com/abstract=4125881 or http://dx.doi.org/10.2139/ssrn.4125881.

      ].
      Aseptic loosening of the tibial component was the most common cause of early revision. Surgeon volume and cement viscosity were associated with an increased rate of failure. Aseptic failure was not seen in any case where the revision tibia and the newly designed tibial baseplate were used. We believe our findings demonstrate that cement debonding is a potential issue with the original design of the tibial component. The new tibial base plate, ATTUNE S+ was designed in 2017 to enhance tibial fixation. Four cement pockets were added for macro-mechanical fixation combined with a 45-degree undercut geometry to provide a macrolock at the cement-implant interface. Additionally, a microblast surface finish increased the surface roughness (3.0-6.5 Ra) compared to the original design (2.5-3.5 Ra) (Fig. 1). The new design of the tibial component appears to have resolved these issues although our series was not able to demonstrate statistically significant difference in performance between tibial components, likely due to a small sample size.
      The results of this study should be interpreted in light of its strengths and weaknesses. Major strengths of this study are the large sample size and low loss to follow-up. There are several limitations of this study. It was conducted at a single health system with a relatively small number of surgeons with limited number of arthroplasty-fellowship-trained as well as low-volume surgeons for comparison. Excluding patients with less than 2 years of follow-up has the potential to misreport the reported data. The high rate of revision associated with 1 brand of cement, only used by 1 low-volume surgeon, introduces bias into the study. Additionally, we did not perform radiographic assessment to determine additional factors leading to cement-implant interface failure. As a retrospective, comparative, exploratory study testing multiple risk factors, caution should be taken when interpreting statistical significance of these findings, but the evidence suggests that further study of cement type and baseplate design could be warranted.
      We echo the calls of other authors in the search of additional peer-reviewed clinical results on this total knee system. An update on the prospective clinical trials sponsored by the manufacturer (clinicaltrials.gov identifiers NCT01746524 and NCT01754363) will provide much-needed insight into the long-term success of this total knee system in large prospective trials.

      Conclusions

      This study represents the largest nonregistry review of the original ATTUNE total knee system. In this retrospective review at our institution, surgeon case volume and cement viscosity were factors associated with an increased rate of early failure due to tibial baseplate implant-cement interface debonding.

      Conflicts of interest

      Dr. Christopher Damsgaard has served as the principal investigator on project that received institutional research suppoer from Breg Inc. and his wife is an employee of Pfizer. Dr. Graham has served as the principal investigator on projects that received institutional research support from Pfizer Inc., Medtronic Inc., Astra Zeneca, and Purdue Pharma LP. All other authors declare no potential conflicts of interest.
      For full disclosure statements, refer to https://doi.org/10.1016/j.artd.2022.06.012.

      Appendix A. Supplementary Data

      References

        • Kurtz S.M.
        • Ong K.L.
        • Lau E.
        • Bozic K.J.
        Impact of the economic downturn on total joint replacement demand in the United States.
        J Bone Joint Surg Am. 2014; 96: 624-630https://doi.org/10.2106/JBJS.M.00285
        • Sharkey P.F.
        • Lichstein P.M.
        • Shen C.
        • Tokarski A.T.
        • Parvizi J.
        Why are total knee arthroplasties failing today—has anything changed after 10 years?.
        J Arthroplasty. 2014; 29: 1774-1778https://doi.org/10.1016/j.arth.2013.07.024
        • Delanois R.E.
        • Mistry J.B.
        • Gwam C.U.
        • Mohamed N.S.
        • Choksi U.S.
        • Mont M.A.
        Current epidemiology of revision total knee arthroplasty in the United States.
        J Arthroplasty. 2017; 32: 2663-2668https://doi.org/10.1016/j.arth.2017.03.066
        • Bonutti P.M.
        • Khlopas A.
        • Chughtai M.
        • Cole C.
        • Gwam C.U.
        • Harwin S.F.
        • et al.
        Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface.
        J Knee Surg. 2017; 30: 435-439https://doi.org/10.1055/s-0037-1603756
        • Nakama G.Y.
        • Peccin M.S.
        • Almeida G.J.
        • Neto O.D.A.L.
        • Queiroz A.A.
        • Navarro R.D.
        Cemented, cementless or hybrid fixation options in total knee arthroplasty for osteoarthritis and other non-traumatic diseases.
        Cochrane Database Syst Rev. 2012; 10: CD006193https://doi.org/10.1002/14651858.CD006193.pub2
        • Dutiel S.
        • Atkinson A.
        Confidence in the ATTUNE® knee is driven by real world scientific evidence.
        ([accessed 08.03.21])
      1. The Attune primary total knee replacement system value dossier. Depuy Synthes 2012. Pg. 77.
        ([accessed 08.03.21])
        • Ranawat C.S.
        • White P.B.
        • West S.
        • Ranawat A.S.
        Clinical and radiographic results of Attune and PFC sigma knee designs at 2-year follow-up: a prospective matched-pair analysis.
        J Arthroplasty. 2017; 32: 431-436https://doi.org/10.1016/j.arth.2016.07.021
        • Turgeon T.R.
        • Gascoyne T.C.
        • Laende E.K.
        • Dunbar M.J.
        • Bohm E.R.
        • Richardson C.G.
        The assessment of the stability of the tibial component of a novel knee arthroplasty system using radiostereometric analysis.
        Bone Joint J. 2018; 100-B: 1579-1584https://doi.org/10.1302/0301-620X.100B12.BJJ-2018-0566.R1
        • Cerquiglini A.
        • Henckel J.
        • Hothi H.
        • Allen P.
        • Lewis J.
        • Eskelinen A.
        • et al.
        Analysis of the Attune tibial tray backside: a comparative retrieval study.
        Bone Joint Res. 2019; 8: 136-145https://doi.org/10.1302/2046-3758.83.BJJ-2018-0102.R2
      2. ATTUNE S+ Technology Overview. © DePuy Synthes 20172019.
        ([accessed 02.09.22])
        • Bonutti P.
        Response to: confidence in the ATTUNE Knee is Driven by Real-World Scientific Evidence: response to Bonutti et al. Article.
        J Knee Surg. 2018; 31: 811-814https://doi.org/10.1055/s-0037-1608945
        • National Joint Registry for England, Wales
        Northern Ireland and the Isle of Man. Implant summary report for DePuy ATTUNE CR and ATTUNE PS. NJR Database extracted April 7, 2017:1–17.
        ([accessed 08.03.21])
        • Kopinski J.E.
        • Aggarwal A.
        • Nunley R.M.
        • Barrack R.L.
        • Nam D.
        Failure at the tibial cement-implant interface with the use of high-viscosity cement in total knee arthroplasty.
        J Arthroplasty. 2016; 31: 2579-2582https://doi.org/10.1016/j.arth.2016.03.063
        • Dwyer K.
        • Jones R.E.
        • Lesko J.
        • Leopold J.
        • Diaz R.
        ATTUNE knee system early performance: minimum two year clinical results.
        ([accessed 08.03.21])
        • Kaptein B.L.
        • den Hollander P.
        • Thomassen B.
        • Fiocco M.
        • Nelissen R.G.
        A randomized controlled trial comparing tibial migration of the ATTUNE cemented cruciate-retaining knee prosthesis with the PFC-sigma design.
        Bone Joint J. 2020; 102: 1158-1166https://doi.org/10.1302/0301-620X.102B9.BJJ-2020-0096.R1
        • Sadauskas A.
        • Engh III, C.
        • Mehta M.
        • Levine B.
        Implant interface debonding after total knee arthroplasty: a new cause for concern?.
        Arthroplasty Today. 2020; 6: 972-975https://doi.org/10.1016/j.artd.2020.07.043
        • Lachiewicz P.F.
        • Vovos T.J.
        • Steele J.R.
        • Wellman S.S.
        Cemented ATTUNE® fixed- bearing modular posterior-stabilized knee arthroplasty has an unexpectedly high rate of revision.
        in: Orthopaedic Proceedings. 102. The British Editorial Society of Bone & Joint Surgery, London2020: 67
        • Lau R.L.
        • Perruccio A.V.
        • Gandhi R.
        • Mahomed N.N.
        The role of surgeon volume on patient outcome in total knee arthroplasty: a systematic review of the literature.
        BMC Musculoskelet Disord. 2012; 13: 250
        • National Joint Registry for England, Wales
        Northern Ireland and the Isle of Man, 13th annual report 2016; Table 3.28.
        ([accessed 08.03.21])
      3. Australian orthopaedic association national joint replacement registry annual report 2016. Tables KT9 and KT22.
        ([accessed 08.03.21])
        • Giaretta S.
        • Berti M.
        • Micheloni G.M.
        • Ceccato A.
        • Marangoni F.
        • Momoli A.
        Early experience with the ATTUNE total knee replacement system.
        Acta Biomed. 2019; 90: 98-103https://doi.org/10.23750/abm.v90i12-S.8997
        • Molloy I.B.
        • Keeney B.J.
        • Sparks M.B.
        • Paddock N.G.
        • Koenig K.M.
        • Moschetti W.E.
        • et al.
        Short term patient outcomes after total knee arthroplasty: does the implant matter?.
        Knee. 2019; 26: 687-699https://doi.org/10.1016/j.knee.2019.01.018
        • Kelly B.C.
        • Owen J.R.
        • Shah S.C.
        • Johnson A.J.
        • Golladay G.J.
        • Kates S.L.
        A biomechanical comparison of the effect of baseplate design and bone marrow fat infiltration on tibial baseplate pullout strength.
        J Arthroplasty. 2021; 36: 356-361https://doi.org/10.1016/j.arth.2020.07.049
        • Staats K.
        • Wannmacher T.
        • Weihs V.
        • Koller U.
        • Kubista B.
        • Windhager R.
        Modern cemented total knee arthroplasty design shows a higher incidence of radiolucent lines compared to its predecessor.
        Knee Surg Sports Traumatol Arthrosc. 2019; 27: 1148-1155https://doi.org/10.1007/s00167-018-5130-0
        • Hazelwood K.J.
        • O’Rourke M.
        • Stamos V.P.
        • McMillan R.D.
        • Beigler D.
        • Robb III, W.J.
        Case series report: early cement–implant interface fixation failure in total knee replacement.
        Knee. 2015; 22: 424-428https://doi.org/10.1016/j.knee.2015.02.016
        • Buller L.T.
        • Rao V.
        • Chiu Y.F.
        • Nam D.
        • McLawhorn A.S.
        Primary total knee arthroplasty performed using high-viscosity cement is associated with higher odds of revision for aseptic loosening.
        J Arthroplasty. 2020; 35: S182-S189https://doi.org/10.1016/j.arth.2019.08.023
        • Kazarian G.S.
        • Lawrie C.M.
        • Barrack T.N.
        • Donaldson M.J.
        • Miller G.M.
        • Haddad F.S.
        • et al.
        The impact of surgeon volume and training status on implant alignment in total knee arthroplasty.
        J Bone Joint Surg Am. 2019; 101: 1713-1723https://doi.org/10.2106/JBJS.18.01205
        • Namba R.S.
        • Cafri G.
        • Khatod M.
        • Inacio M.C.
        • Brox T.W.
        • Paxton E.W.
        Risk factors for total knee arthroplasty aseptic revision.
        J Arthroplasty. 2013; 28: 122-127https://doi.org/10.1016/j.arth.2013.04.050
        • Martin R.
        • Mason B.
        • Wronski P.
        • Schilkowsky R.
        • Orfanos A.
        • Fehring T.
        • et al.
        Motion during cementing significantly decreases tibial implant fixation strength.
        J Arthroplasty. 2022; 37: S12-S18https://doi.org/10.1016/j.arth.2022.02.091
        • Abdel M.P.
        • Bonadurer III, G.F.
        • Jennings M.T.
        • Hanssen A.D.
        Increased aseptic tibial failures in patients with a BMI > 35 and well-aligned total knee arthroplasties.
        J Arthroplasty. 2015; 30: 2181-2184
        • Pfefferle K.J.
        • Gil K.M.
        • Fening S.D.
        • Dilisio M.F.
        Validation study of a pooled electronic healthcare database: the effect of obesity on the revision rate of total knee arthroplasty.
        Eur J Orthop Surg Traumatol. 2014; 24: 1625-1628
        • Dowsey M.
        • Choong P.M.
        Obese diabetic patients are at substantial risk for deep infection after primary TKA.
        Clin Orthop Relat Res. 2009; 467: 1577-1581
        • Stickles B.
        • Phillips L.
        • Brox W.T.
        • Owens B.
        • Lanzer W.L.
        Defining the relationship between obesity and total joint arthroplasty.
        Obes Res. 2001; 9: 219-223
      4. Cox Z., Engstrom S., Shinar A., Polkowski G., Mason B., Martin R., et al. Is cement mantle thickness a primary cause of aseptic loosening following primary total knee arthroplasty? TheKNE-D-22-00258. SSRN: https://ssrn.com/abstract=4125881 or http://dx.doi.org/10.2139/ssrn.4125881.