2 Results
- Surgical techniqueOpen Access
Technical Note for Transfemoral Implantation of Tapered Revision Stems. The Advantage to Stay Short
Arthroplasty TodayVol. 9p16–20Published online: April 23, 2021- Bernd Fink
Cited in Scopus: 1The aim was to test the hypothesis that during transfemoral implantation of a conical revision stem, the fixation of the stem at the distal tip leads to a low rate of periprosthetic fractures. - Surgical techniqueOpen Access
What can the surgeon do to reduce the risk of junction breakage in modular revision stems?
Arthroplasty TodayVol. 4Issue 3p306–309Published online: April 20, 2018- Bernd Fink
Cited in Scopus: 17Modular revision stems are very common in hip arthroplasty, but junction fracture remains a known failure mechanism. A review of the literature with description of cases with junction breakage of modular revision stems showed that in all 24 analyzed cases, there was a common finding: the combination of an effective osteointegration of the distal component and missing medial bone support of the proximal component. The result was a bending stress point of the stem construction in the region of the junction.